Evaluation of a novel polymer-free sirolimus-eluting stent in a porcine model

CHEN Ming,ZHENG Bo,PENG Hongyu
2009-01-01
Abstract:Objective To evaluate the efficacy and safty of Polymer-free sirolimus eluting-stents(PFSES) in preventing restenosis in porcine coronary model.Methods Bare metal stents(BMS)(n=13),polymer-free bare metal stents(PFBMS)(n=13),polymeric sirolimus-eluting stents(PSES)(n=13) and PFSES(n=13) were implanted in left anterior descending coronary arteries(n=26) and left circumflex coronary(n=26) of 27 swines randomly.Coronary angiography was performed 28 days and 90 days after the procedure.Twelve animals were sacrificed for histomorphologic analysis after 90 days.Results The 28-day(n=24) and 90-day(n=12) outcome by quantitative coronary angiography(QCA) showed significant reduction in luminal loss(LL) in PFSES(LL:0.69±0.49 mm vs 1.27±0.36 mm,P=0.041;0.77±0.44 mm vs 1.33±0.29 mm,P<0.01)compared with BMS.By histomorphomeric analysis,with the similar injury sores,there was also a corresponding significant reduction in neointimal tissue proliferation in PFSES compared with BMS control(average neointimal area: 2.412±1.149 mm2 vs 4.475±1.345 mm2,P<0.05).High magnification histomorlogic examination revealed similar inflammation score between PFSES and BMS,both of which were lower than PSES.Conclusion Polymer-free sirolimus-eluting stents can reduce neointimal hyperplasia with good biocompatibility in porcine coronary model.
What problem does this paper attempt to address?