ADSORPTIVE STRIPPING VOLTAMMETRIC STUDY OF SCANDIUM–ALIZARIN COMPLEXAN COMPLEX AT A CARBON PASTE ELECTRODE

Junan Li,Fenyun Yi,Dongsheng Shen,Junjie Fei
DOI: https://doi.org/10.1081/AL-120006672
2007-01-01
Analytical Letters
Abstract:A highly selective, sensitive, rapid and economic adsorptive stripping voltammetric method has been. developed for the determination of ultra trace amount of scandium at a carbon paste electrode (CPE) in the current paper. The Sc(III)-alizarin complexan (ALC) complex is adsorbed at the electrode in a mixed buffer solution(pH 5.4) containing 0.04 M acetate and 0.016 M biphthalate, yielding one peak corresponding to the reduction of ALC in the complex at -0.60V (vs. SCE). The second-order derivative linear scan voltammograms of the complex are recorded by use of model JP-303 polarographic analyzer from -0.10 to -1.10 V. A linear relationship between the second order derivative peak current and Sc(III) concentration in the range of 1.0 x 10(-9) to 6.0 x 10(-7) M is observed. The detection limit for Sc(III) is 6.0 x 10(-10) M. A relative standard deviation of 4.34% is given for ten determinations of 6.0 x 10(-8) M Sc(III). The method has been applied to the determination of the trace Sc(III) in the mineral samples with satisfactory results.