Study on Detection of Peroxide Value in Edible Oils Using FTIR Spectral Reconstitution Technique

Yu Xiuzhu,Du Shuangkui,Li Zhixi,Zhang Jingya
DOI: https://doi.org/10.3969/j.issn.1009-7848.2011.07.026
2011-01-01
Abstract:The paper took edible oils,triphenylphosphine(TPP),triphenylphosphine oxide(TPPO) and odorless mineral spirit(OMS) as experiment material,adopted Fourier transform infrared spectral reconstitution(SR-FTIR)technique to detect peroxide value(PV) in edible oils,basing on the PV determination is the rapid reaction of TPP with the hydroperoxides present in an oil to produce TPPO,which exhibits a readily measurable absorption band at 542 cm-1.The result showed that calibration standards,prepared by gravimetric addition of TPPO to a peroxide-free oil,were handled in the same manner,and a linear calibration equation relating the concentration of TPPO(expressed as the equivalent PV) to the absorbance of TPPO at 542 cm-1 relative to a baseline at 530 cm-1 in the reconstituted spectra was obtained,with a regression SD of ±0.067 mmol/kg oil.PV determinations on two sets of validation samples,spanning PV ranges of 0~1.5 and 1.5~10 mmol/kg oil,were carried out in parallel by the iodimetry and SR-FTIR-based FTIR procedures,and comparison of the results of duplicate analyses by the two methods indicated that the latter was more reproducible and slightly more sensitive.The SR-FTIR-based PV method,when implemented on an autosampler-equipped FTIR system,allowed for the automated analysis of 90 samples per hour.