Automatic Determination of Acid Value and Peroxide Value of Edible Oils by Near-infrared Spectroscopy

Yu Xiuzhu,Zhang Jingya,Li Qinghua,Xu Chunjin,Gao Jinming
DOI: https://doi.org/10.6041/j.issn.1000-1298.2012.09.028
2012-01-01
Abstract:For automatic determination of acid value and peroxide value of edible oils, using edible oils as raw materials, calibrations were developed and validated for determination of acid value and peroxide value by near-infrared spectrometer with continuous samples injection cell. The correlation coefficient R and the root mean square errors of prediction (RMSEP) of the calibration for acid value were 0.9873 and 0.114 mg/g in range of 5500-4600 cm -1 with baseline correction point at 6524 cm -1 and 4823 cm -1 and standard normal variant (SNV). The R and RMSEP for peroxide value of calibration were 0.9958 and 0.90 mmol/kg in range of 6050-4450 cm -1 with first derivative and Norris filter. The results of blind samples validation were good, the calibrations were workable, throughput of 90 samples can be achieved per hour.
What problem does this paper attempt to address?