Rapid Detection Of Gly, L-Arg, L-Asp Using An Saw-Impedance Sensor
Ronghui Wang,Defang Tong,Wei Wei,Lihua Nie,Shouzhuo Yao
DOI: https://doi.org/10.1080/10739149709351447
1997-01-01
Instrumentation Science & Technology
Abstract:The surface acoustic wave (SAW)-impedance sensor, which is composed of a 61 MHz SAW resonator and a pair of parallel platinum electrodes, has been successfully applied to rapid detection of glycine (Gly), L-arginine (L-Arg) and L-aspartic acid (L-Asp) and is reported in this paper. Two methods are described - the calibration curve method and the frequencimetric titration method. The effect of formaldehyde solution on the frequency shift is discussed. For the calibration curve method: in unpretreated formaldehyde medium, the detection limit of Gly, L-Asp, L-Arg is 1.40x10(-5)M, 1.48x10(-6)M, 1.04x10(-5)M and the frequency shift is linear up to 1.83x10(-3)M, 1.60x10(-3)M, 1.37x10(-3)M; in pretreated formaldehyde medium, the above values are 7.85x10(-6)M, 1.11x10(-6)M, 2.30x10(-6)M and 1.53x10(-3)M, 1.51x10(-3)M, 0.96x10(-3)M, respectively. Using the frequencimetric titration method, the lowest titratable concentration of Gly, L-Asp, L-Arg is 1.14x10(-5)M, 1.07x10(-5)M, 1.70x10(-5)M. A comparison study between the SAW method and some other previously reported techniques is discussed in this paper.