Abstract:Objective To evaluate clinical outcomes of a novel aiming device for the placement of cannulated screws guide pins in the treatment of femoral neck fractures.Methods We made a novel aiming device and applied it to clinical use.Between January 2010 and November 2011,there were 40 femoral neck fracture patients who were included into the study.All 40 patients were divided into two groups randomly.The aiming device was used during the surgery for patients of experimental group while the aiming device was not allowed to use during the surgery for patients of control group.Results There were no statistically significant difference in surgery time and bleed volume between two groups(P>0.05).The number of guide pin placement were significantly fewer in experimental group than control group(P<0.05),meanwhile,the angle between the first cannulated screw and the middle line of femoral neck in coronal plane,the angle between the first cannulated screw and the middle line of femoral neck in sagittal plane and the distance between cannulated screws and femoral calcar were significantly smaller in experimental group than control group(P<0.05).Conclusion The aiming device,which is simple in structure and easy to use,could help surgeons to place cannulated screws guide pins,and it is valuable to apply it to clinical use widely.