Effect of Respiratory Motion on the GTV Contoured with PET Image

CHEN Shao-qing,YU Jin-ming,YIN Yong,FU Zheng,XING Li-gang,LI Shan-shan,GONG He-yi,LIU Tong-hai,TENG Xue-peng,LU Jie
DOI: https://doi.org/10.3969/j.issn.1673-5269.2007.07.018
2007-01-01
Abstract:OBJECTIVE:To evaluate the effect of respiratory motion on GTV contoured with PET image, using a 2D respiratory motion phantom system. METHODS: Twenty-eight kinds of respiratory motions were simulated by using a respiratory motion phantom system. The phantom was scanned by PET-CT machine under different motion conditions, then the PET images were obtained. GTV was contoured by using PET images. RESULTS: GTV under static condition (GTV0) was (7.90 ± 0.21) cm3. GTV with motion frequencies of 16 per minute, 18 per minute and 20 per minute were (11.56 ± 2.62) cm3 (GTV16), (12.51 ± 3.57) cm3 (GTV18) and 11.86 ± 3.27 cm3 (GTV20), respectively. The differences among GTV0, GTV16, GTV18 and GTV20 were significant (all P0.01). GTV with motion amplitudes of 1.0 cm, 1.5 and 2.0 cm on z axis were (11.44±3.68) cm3 (GTVz-1.0), 11.64±3.47 cm3(GTVz-1.5) and (12.83±2.02) cm3 (GTVz-2.0), respectively. GTV with motion amplitudes of 0.5 cm, 1.0 cm and 1.5 cm on x axis were (9.68 ± 2.32) cm3 (GTVx-0.5), (14.41 ± 3.19) cm3 (GTVx-1.0) and (11.83 ± 1.92) cm3 (GTVx-1.5) respectively. The differences of GTV on z axis and x axis were significant (all P0.01). CONCLUSION: GTV contoured with PET image will be severely effected by respiratory motion which will impact on radiation treatment planning. The gating technique should be used in order to decrease this kind of effect when GTV is contoured with PET image.