Accuracy of magnetic resonance imaging in the localization of stereotactic radiotherapy

DIAO Huan-rong,LIU Xiao-lan,JIANG Sha,LIU Ying,CAO Yong-zhen
2006-01-01
Abstract:Objective: To measure the errors of magnetic resonance imaging(MRI) in the localization of stereotactic radiotherapy(SRT),and to confirm whether MRI can be used in the localization alone.Methods: A specially made phantom was used to acquire MRI images.In the treatment planning system,the distances between targets in x,y,and z directions were measured and compared with true values.The absolute values of the differences between the measured and true values were considered as the errors of the localization system.Results: The errors of adjacent points were in the range of 0.2 to 1.4 mm and 0.3 to 1.5 mm in x and y directions,respectively,and the average was 0.9 mm in both x and y directions.The errors of opposite points were in the range of 0.3 to 1.7 mm and 0.4 to 1.8 mm in x and y directions,respectively,and the average was 0.9 mm in both x and y directions.There was no significant difference among 6 marked planes in both x and y directions(P > 0.05).Errors in z direction were in the range of 0.3 to 1.9 mm,and the average was 1.0 mm.Conclusion: MRI can be used alone in the localization of most brain tumors and head and neck tumors,but it is necessary to perform a comprehensive check when MRI is used in SRT localization.