Application of low atrial septal pacing.

SU Yan-gang,WANG Wei,BAI Jin,GE Jun-bo
2009-01-01
Abstract:Objective To evaluate the feasibility and safety of low atrial septal(LAS,the region of Koch triangle) pacing compared with the routine right atrial auricle(RAA) pacing.Methods Sixty patients who applied DDD pacemaker were randomized equally into RAA and LAS pacing groups.In the LAS pacing group,active fixation lead was used to locate to the site of RAA followed by screwing in the site of LAS.The difference in the pacing parameters,lead implantation success rate,X ray exposure time and dislocation rate during and after operation were investigated.Results ①Pacing parameters:there were no difference in pacing threshold and impedance between two pacing sites.However,compared with RAA pacing site,the amplitude of P wave was higher(3.8±0.7 mV vs 2.2±0.8 mV) and width of pacing P wave was narrower (88±18 ms vs 154±37 ms) in LAS pacing site.②Lead implantation success rate and X ray exposure time : the success rate of LAS was a little lower than RAA group(90% vs 100%,P>0.05),and the X ray exposure time in LAS site was much longer than RAA site(128±45 s vs 12±4 s).③Lead dislocation: during operation the lead dislocation of LAS and RAA site were 33.3% and 0%(P<0.05),respectively.Two cases of LAS group were found atrial lead dislocation during follow up,while no dislocation was in RAA group. Conclusion LAS pacing is feasible and markedly shortens biatrial activation times though the fixation to the LAS is more complicated compared with RAA pacing.[Chinese Journal of Cardiac Pacing and Electrophysiology,2009,23(1):19-22]