Comparison of effects of pacing atrial septum and right atrial appendage on atrial activation

C. Shao-liang
Abstract:Objective To compare the effects of right atrial appendage(RAA) pacing and atrial septal(AS) pacing on the atrial activation.Methods RAA and AS pacing were performed in patients(42 cases and 27 cases) who were received DDD artificial pacemaker implant.A 12-lead surface electrogram was recorded.P wave duration maximum(Pmax) and dispersion(Pd) was measured in these patients.Results Pmax and Pd was prolonged during RAA,Pmax post-opration(138±23) ms vs pre-operation(127±16) ms,Pd post-operation((19.4)±(9.0)) ms vs pre-operation((13.5)±(7.3)) ms pacing and shortened during AS pacing,Pmax post-operation(122±12) ms vs pre-operation(133±17) ms,Pd post-operation((11.1)±(6.6)) ms vs pre-operation((15.5)±(8.0)) ms(P(0.05)).Conclusion Compared to RAA pacing,AS pacing markedly reduces atrial activation duration and atrial dispersion,which is in favor of preventing atrial fibrillation.