The Hemolysis Research of the National Implantable Ventricular Assist Device

LI Tie-yan,YE Liang,ZHANG Zhi-guo,FAN Hui-min,LIU Zhong-min
DOI: https://doi.org/10.3969/j.issn.1008-0392.2011.02.011
2011-01-01
Abstract:Objective To evaluate the hemolysis of a home-made implantable ventricular assist device(NIVAD).Methods Two thousand ml calf blood were anticoagulated at the proportion of 1:9 with 3.8%sodium citrate.Then adjusted the flow as 4 L/min and the outflow pressure 13.3 kPa through the damper valve and rotational speed of NIVAD,to achieve NIVAD working. Blood samples were taken before and also 0.5,1,1.5...4h after working.Free hemoglobin (FHB) and hematocrit(HCT) were measured.During the same time,the unused calf blood was taken as control.Results The FHB of experimental blood is higher than that of the control.The normal index of hemolysis(NIH) is(0.064 965±0.014 846) mg/L.Conclusion Shearing force resulting from the high speed rotation of NIVAD does a certain degree of destruction to RBC,but the NIH is in criterion range of hemolysis.