Spectroscopic Measurements of Impurity Spectra on the EAST Tokamak

Jia Fu,Yingying Li,SHI Yuejiang,Fudi Wang,Wei Zhang,Bo Lv,HUANG Juan,WAN Baonian,ZHOU Qian
DOI: https://doi.org/10.1088/1009-0630/14/12/03
2012-01-01
Abstract:Ultraviolet (UV) and visible impurity spectra (200 similar to 1750 nm) are commonly used to study plasma and wall interactions in magnetic fusion plasmas. Two optical multi-channel analysis (OMA) systems have been installed for the UV-visible spectrum measurement on EAST. These two OMA systems are both equipped with the Czerny-Turner (C-T) type spectrometer. The upper vacuum vessel and inner divertor baffle can be viewed simultaneously through two optical lenses. The OMA1 system is mainly used for multi-impurity lines radiation measurement. A 280 nm wavelength range can be covered by a 300 mm focal length spectrometer equipped with a 300 grooves/mm grating. The D alpha/H alpha line shapes can be resolved by the OMA2 system. The focal length is 750 mm The spectral resolution can be up to 0.01 nm using a 1800 grooves/mm grating. The impurity behaviour and hydrogen ratio evolution after boroniztion, lithium coating, and siliconization are compared. Lithium coating has shown beneficial effects on the reduction of edge recycling and low Z impurity (C, O) influx. The impurity expelling effect of the divertor configuration is also briefly discussed through multi-channels observation of OMA1 system.