Study on Temperature Compensation for Matched Grating Interrogation System

翟玉锋,张龙,李飞,于清华,刘勇,王安
2008-01-01
Abstract:A novel matched grating interrogation system with temperature compensation was proposed to eliminate the influence on measurement precision of the sensing FBG causured by cross sensitivity.The principle of the temperature compensation is that the matching FBG's temperature is balanced with the shift of FBG sensor's temperature.An additional pair of matched gratings and a constant temperature controller based on the semiconductor cooler which was clos-loop self-controlled by the controller MAX1968EUI were employed in this system.An cheap micro-load cell as the output element of matched grating interrogation system was also proposed,different from PZT as the output element of the custom dynmatic matched grating interrogation system.The experiment results demonstrated that the inaccuracy of strain measurement of this system was less than 2%,under the temperature variation.