High Sensitive Detection of Trace Heavy Metals in Water by Laser-induced Breakdown Spectroscopy

Fang ZHAO,Qian ZHANG,Wei XIONG,Jing-bao RONG,Run-hua LI
DOI: https://doi.org/10.3969/j.issn.1003-6504.2010.03.030
IF: 11.4
2010-01-01
Environmental Science & Technology
Abstract:Trace toxic heavy metal ions in natural water were electrically deposited on the surface of an aluminum rod with 99.999% purity and then quantitatively analyzed by laser-induced breakdown spectroscopy (LIBS). The interference of gas bubbles generated in the process of electrolysis to electrical-deposition of heavy metal ions was effectively reduced by utilizing a rotating cathode. Pb and Cd were analyzed in experiments and calibration curves of them were established under the condition of 7.5V of deposition voltage and 20min of deposition time. Limits of detection achieved were 500ng/L for Pb and 830ng/L for Cd. The method can provide a new analytical technique for monitoring heavy metal pollutions of water with high detection sensitivity and possibility of multi-elements simultaneous analysis.
What problem does this paper attempt to address?