Development of Micro Hemoglobin Sensor Array Based on MEMS

SHI Jian,YAN Guo-zheng,FANG Yun
2006-01-01
Abstract:Based on the technology of micro-process and micro-fabrication,a micro hemoglobin sensor array is developed.This one-chip sensor has tiny nonlinearity of 3.67%,low detecting limit of 0.1 mg/mL and high sensitivity of 70 μA in 1 mg/mL hemoglobin solution.The deviation is not more than 10% caused by one unit change of pH value on acid condition,futhermore in the case of neutral or alkali condition the deviation is undistinguished.The deviation caused by temperature change is only 0.2 μA/℃.The high linearity,stable pH dependence and temperature dependence bring the one-chip sensor large prospect in application of medical field such as hemoglobin's measurement and gastrointestinal bleeding's check.