Reliability and sensitivity of an instrument for measuring the midfoot passive mechanical properties
Fabrício Anicio Magalhães,Thales Rezende Souza,Vanessa Lara Araújo,Renan Alves Resende,Antônio Eustáquio de Melo Pertence,Mauro Heleno Chagas,Sérgio Teixeira Fonseca
DOI: https://doi.org/10.1016/j.jbiomech.2020.109735
2020-05-07
Abstract:To assess the test - retest reliability and sensitivity of an instrument developed to measure the passive mechanical properties of the midfoot joint complex (MFJC), nine female and three male healthy young adults were evaluated in two different days by two experienced examiners. After proper participant positioning on the instrument, the left forefoot was passively moved from eversion (20°) to inversion (45°) at 2°/s, while the rearfoot and shank were kept immobile. From the instrument's sensing units (torque meter and potentiometer at 100 Hz), passive torque and angle data were registered. Passive stiffness was calculated as the instantaneous slope of the torque vs. angle curve. Therefore, the variables analyzed were resting angle, passive torques and stiffnesses during inversion and eversion through the intraclass correlation coefficient (ICC3,3), standard error of measurement (SEM) and minimal detectable change (MDC95). For the resting angle, ICC3,3 ranged from 0.85 to 0.91, SEM ranged from 1.54° to 1.95° and MDC95 ranged from 4.26° to 5.41°. For the torques, ICC3,3 ranged from 0.85 to 0.97, SEM ranged from 0.09Nm to 0.42Nm and MDC95 ranged from 0.26Nm to 1.16Nm. Finally, for the stiffnesses, ICC3,3 ranged from 0.79 to 0.98, SEM ranged from 0.01Nm/° to 0.04Nm/° and MDC95 ranged from 0.01Nm/° to 0.10 Nm/°. It can be concluded that most of the measurements presented good to excellent reliability with low measurement error. Hence, clinicians and researchers may benefit from the reliable and stable measures provided by the Foot Torsimeter when assessing patients and planning interventions.