Technical note: Characterization of a multi‐point scintillation dosimetry research platform for a low‐field MR‐Linac

Jennie Crosby,Chase Ruff,Ken Gregg,Jonathan Turcotte,Carri Glide‐Hurst
DOI: https://doi.org/10.1002/mp.17192
IF: 4.506
2024-06-07
Medical Physics
Abstract:Background MRI‐guided radiation therapy (MRgRT) requires unique quality assurance equipment to address MR‐compatibility needs, minimize electron return effect, handle complex dose distributions, and evaluate real‐time dosimetry for gating. Plastic scintillation detectors (PSDs) are an attractive option to address these needs. Purpose To perform a comprehensive characterization of a multi‐probe PSD system in a low‐field 0.35 T MR‐linac, including detector response assessment and gating performance. Methods A four‐channel PSD system (HYPERSCINT RP‐200) was assembled. A single channel was used to evaluate repeatability, percent depth dose (PDD), detector response as a function of orientation with respect to the magnetic field, and intersession variability. All four channels were used to evaluate repeatability, linearity, and output factors. The four PSDs were integrated into an MR‐compatible motion phantom at isocenter and in gradient regions. Experiments were conducted to evaluate gating performance and tracking efficacy. Results For repeatability, the maximum standard deviation of repeated measurements was 0.13% (single PSD). Comparing the PSD to reference data, PDD had a maximum difference of 1.12% (10 cm depth, 6.64 × 6.64 cm2). Percent differences for rotated detector setups were negligible (