Fully automated workflow for integrated sample digestion and Evotip loading enabling high-throughput clinical proteomics

Anders H. Kverneland,Florian Harking,Joel Mario Vej-Nielsen,Magnus Huusfeldt,Dorte B. Bekker-Jensen,Inge Marie Svane,Nicolai Bache,Jesper V. Olsen
DOI: https://doi.org/10.1016/j.mcpro.2024.100790
IF: 7.381
2024-05-23
Molecular & Cellular Proteomics
Abstract:Protein identification and quantification is an important tool for biomarker discovery. With the increased sensitivity and speed of modern mass spectrometers, sample-preparation remains a bottleneck for studying large cohorts. To address this issue, we prepared and evaluated a simple and efficient workflow on the Opentrons OT-2 (OT-2) robot that combines sample digestion, cleanup and loading on Evotips in a fully automated manner, allowing the processing of up to 192 samples in 6 hours. Analysis...
biochemical research methods
What problem does this paper attempt to address?