A novel hollow-electrode glow discharge mass spectrometry for quantitative analysis of protein content in food

Min Ren,Yingqi Wan,Jiwen Chen
DOI: https://doi.org/10.1039/d4ay01022a
IF: 3.1
2024-07-14
Analytical Methods
Abstract:Protein content in food is an important indicator of nutritional value and food safety.Therefore it is great significance to accurately detect protein content in food. In this paper, a combustion furnace and a novel hollow-electrode glow discharge ion source-quadrupole mass spectrometry(HGD-MS) were designed. Use them constructing a "combustion furnace + mass spectrometry" experimental platform to detect the protein content in food. Five food standard samples were selected for analysis. The food samples were combusted in the combustion furnace at a high temperature (1300°C) in an oxygen-rich environment. The gas products were passed into the novel hollow electrode glow discharge ion source-quadrupole mass spectrometry. A standard curve of y = 64365x - 29953 ,R2=0.9997 was plotted by detecting the NO+ ion intensity, relative standard deviation (RSD) of 1.8% to 5.7%.Using the same method, food sample No. 6 and 7 were combusted and NO+ ion intensity was measured to verify the accuracy of the quantitation curve. Subsequently, the protein content was determined using a nitrogen-to-protein conversion factor of 6.25. This method provides a rapid, accurate, and environmentally friendly approach for determining protein content in food
spectroscopy,chemistry, analytical,food science & technology