Concurrent Validity and Relative Reliability of the RunScribeTM System for the Assessment of Spatiotemporal Gait Parameters During Walking

Andrés Ráfales-Perucha,Elisa Bravo-Viñuales,Alejandro Molina-Molina,Antonio Cartón-Llorente,Silvia Cardiel-Sánchez,Luis E. Roche-Seruendo
DOI: https://doi.org/10.3390/s24237825
IF: 3.9
2024-12-07
Sensors
Abstract:The evaluation of gait biomechanics using portable inertial measurement units (IMUs) offers real-time feedback and has become a crucial tool for detecting gait disorders. However, many of these devices have not yet been fully validated. The aim of this study was to assess the concurrent validity and relative reliability of the RunScribeTM system for measuring spatiotemporal gait parameters during walking. A total of 460 participants (age: 36 ± 13 years; height: 173 ± 9 cm; body mass: 70 ± 13 kg) were asked to walk on a treadmill at 5 km·h−1. Spatiotemporal parameters of step frequency (SF), step length (SL), step time (ST), contact time (CT), swing time (SwT), stride time (StT), stride length (StL) and normalized stride length (StL%) were measured through RunScribeTM and OptoGaitTM systems. Bland–Altman analysis indicated small systematic biases and random errors for all variables. Pearson correlation analysis showed strong correlations (0.70–0.94) between systems. The intraclass correlation coefficient supports these results, except for contact time (ICC = 0.64) and swing time (ICC = 0.34). The paired t-test showed small differences in SL, StL and StL% (≤0.25) and large in CT and SwT (1.2 and 2.2, respectively), with no differences for the rest of the variables. This study confirms the accuracy of the RunScribeTM system for assessing spatiotemporal parameters during walking, potentially reducing the barriers to continuous gait monitoring and early detection of gait issues.
engineering, electrical & electronic,chemistry, analytical,instruments & instrumentation
What problem does this paper attempt to address?