The Tibial Tunnel Drilling Angles of 60° Provided a Lower Ultimate Load to Failure on a Single Bundle Posterior Cruciate Ligament Graft Using Interference Screw Fixation Compared to 30°/45°
Xiaohui Zhang,Fei Teng,Bin Geng,Fan Lu,Zhongcheng Liu,Laiwei Guo,Hua Han,Meng Wu,Yayi Xia,Yuanjun Teng
DOI: https://doi.org/10.1007/s00167-023-07428-6
2023-01-01
Knee Surgery Sports Traumatology Arthroscopy
Abstract:To biomechanically compare the initial fixation strength of grafts among three tibial tunnel angles (30°/45°/60°) in transtibial posterior cruciate ligament (PCL) reconstruction. A series of transtibial PCL reconstruction models were established with porcine tibias and bovine tendons. Specimens were randomly assigned to three groups according to the angles between the tibial tunnel and the perpendicular line of the tibial shaft: Group A (30°, n = 12), Group B (45°, n = 12), and Group C (60°, n = 12). The area of the tunnel entrance, the segmental bone mineral density (sBMD) of the graft fixation site of the tibia and the maximum insertion torque of the interference screw were measured. Finally, load to failure tests were carried out on the graft-screw-tibia constructs at the same rate. Ultimate load to failure in Group C (335.2 ± 107.5 N) was significantly lower than that in Group A (584.1 ± 127.9 N, P < 0.01) and Group B (521.9 ± 95.9 N, P < 0.01). There were no significant differences between biomechanical properties of Groups A and B (n.s.). The posterior part fractures of the tibial tunnel exit occurred in eight specimens of Group C. In addition, the ultimate load was proven to be related to insertion torque (rho = 0.7, P < 0.01), sBMD (rho = 0.7, P < 0.01), and the area of the tunnel entrance (rho =− 0.4, P = 0.01). The ultimate load to failure was significantly lower in tibial PCL interference screw fixation for tunnels drilled at 60° compared to 30°/45°. In addition, the ultimate load was significantly correlated with insertion torque, sBMD and the area of the tunnel entrance. Given that the load to failure of distal fixation may not be sufficient for early postoperative rehabilitation, a 60° tunnel should not be recommended to drill in tibia during PCL reconstruction.