Pulse oximetry signal loss during hypoxic episodes in preterm infants receiving automated oxygen control

Lukas O. Langanky,Karen B. Kreutzer,Christian F. Poets,Axel R. Franz,Christoph E. Schwarz
DOI: https://doi.org/10.1007/s00431-024-05549-9
2024-04-10
European Journal of Pediatrics
Abstract:The aim of this study was to analyze signal loss (SL) resulting from low signal quality of pulse oximetry-derived hemoglobin oxygen saturation (SpO 2 ) measurements during prolonged hypoxemic episodes (pHE) in very preterm infants receiving automatic oxygen control (AOC). We did a post hoc analysis of a randomized crossover study of AOC, programmed to set FiO 2 to "back-up FiO 2 " during SL. In 24 preterm infants (median (interquartile range)) gestational age 25.3 (24.6 to 25.6) weeks, recording time 12.7 h (12.2 to 13.6 h) per infant, we identified 76 pHEs (median duration 119 s (86 to 180 s)). In 50 (66%) pHEs, SL occurred for a median duration of 51 s (33 to 85 s) and at a median frequency of 2 (1 to 2) SL-periods per pHE. SpO 2 before and after SL was similar (82% (76 to 88%) vs 82% (76 to 87%), p = 0.3)).
pediatrics
What problem does this paper attempt to address?