Abstract:Rosuvastatin calcium is a widely used 3-hydroxy-3-methylglutaryl coenzyme A-reductase inhibitor developed for the treatment of dyslipidemia. To establish a control strategy for the elemental impurities, a new digestion method combined with an inductively coupled plasma-mass spectrometer (ICP-MS) was developed and validated by our team to determine elements Cd, Pb, As, Hg, Co, V, and Ni in rosuvastatin calcium tablets, which digest the sample perfectly even in the presence of a large number of excipients, especially titanium dioxide. The measurement mode was collision cell mode with kinetic energy discrimination (KED). 209Bi+, 115In+, and 89Y+ were chosen as internal standard elements. The recoveries of the limit of quantitation (LOQ) ranged from 90.5% to 106.4%, concentrations of the abovementioned elements in LOQ were 0.25 μg·L−1, 0.25 μg·L−1, 0.75 μg·L−1, 1.5 μg·L−1, 2.5 μg·L−1, 5 μg·L−1, and 8 μg·L−1 , respectively, linear correlation coefficients were above 0.9997, the recoveries in accuracy item ranged from 91.8% to 103.6%, and relative standard deviations (RSDs) of recovery in precision were not more than 1.8%, reflecting a reliable method of high sensitivity, strong anti-interference capacity, and good precision, and that it was suitable for the determination of elemental impurities in drugs.
chemistry, analytical
What problem does this paper attempt to address?
This paper aims to solve the problem of quantitative analysis of elemental impurities in rosuvastatin calcium tablets. Specifically, the research team developed and validated a new digestion method, combined with inductively coupled plasma - mass spectrometry (ICP - MS), for the determination of elemental impurities such as Cd, Pb, As, Hg, Co, V and Ni in rosuvastatin calcium tablets. This method can completely digest samples in the presence of a large amount of excipients, especially for titanium dioxide excipients with extremely high thermal stability and high chemical degradation resistance, and has high sensitivity, strong anti - interference ability and good accuracy. ### Background and Problem of the Paper Rosuvastatin calcium is a widely used 3 - hydroxy - 3 - methylglutaryl - coenzyme A reductase inhibitor, mainly used for the treatment of dyslipidemia, especially in elderly patients with coronary heart disease complicated with hyperlipidemia. Since this type of lipid - lowering drugs needs to be taken for a long time, the control of elemental impurities in drugs is particularly important. According to the requirements of Chapter 232 of the United States Pharmacopeia (USP) and the ICH Q3D guidelines, risk assessment of elements such as Cd, Pb, As, Hg, Co, V and Ni needs to be carried out to ensure that the levels of these elements are within the Permitted Daily Exposures (PDEs). ### Research Methods 1. **Sample Pretreatment**: - Using a microwave digestion system, the digestion conditions were optimized to avoid using sulfuric acid or high temperature (450°C) to reduce the volatilization of mercury. - The digestion parameters include: 5 mL nitric acid, 50 μL hydrofluoric acid, 0.2 mL ICP standard gold, the initial temperature is room temperature, and gradually heated to 190°C and maintained for a period of time. 2. **ICP - MS Detection**: - The most abundant isotopes were selected for quantitative analysis, such as \(^{111}\text{Cd}^+\), \(^{208}\text{Pb}^+\), \(^{75}\text{As}^+\), \(^{202}\text{Hg}^+\), \(^{59}\text{Co}^+\), \(^{51}\text{V}^+\) and \(\text{^{60}Ni}^+\). - The internal standard elements are \(^{209}\text{Bi}^+\), \(^{115}\text{In}^+\) and \(\text{^{89}Y}^+\), which are used to correct the matrix effect. 3. **Method Validation**: - **Linear Range**: Through the test of standard solutions, the linear range and limit of quantification (LOQ) of each element were determined. - **Specificity**: Through the tests of blank solution and sample blank solution, it was ensured that there was no interference. - **Accuracy**: Through the spiked recovery experiment, the accuracy of the method was verified. - **Precision**: Through the repeatability and intermediate precision tests, the precision of the method was verified. ### Results and Discussion - **Linear Range**: The linear ranges and correlation coefficients of all elements meet the requirements, and the correlation coefficients \(R^2\) are all greater than 0.9997. - **Specificity**: No interference signals were found in the blank solution and the sample blank solution. - **Accuracy**: The spiked recovery rates are between 91.8% and 103.6%, which meet the requirements of Chapter 233 of USP. - **Precision**: The relative standard deviations (RSD) of repeatability and intermediate precision do not exceed 2.0%. ### Conclusion This study successfully developed and validated a method suitable for the quantitative analysis of elemental impurities in rosuvastatin calcium tablets. This method can not only effectively process complex samples containing excipients such as microcrystalline cellulose, lactose monohydrate, calcium phosphate, crospovidone, magnesium stearate and titanium dioxide, but also has high sensitivity, anti - interference ability and accuracy. In addition, by optimizing the digestion conditions, the volatilization of mercury is reduced, ensuring the reliability of the analysis results.