Validation of the instant blood pressure smartphone app

Timothy B Plante, Bruno Urrea, Zane T MacFarlane, Roger S Blumenthal, Edgar R Miller, Lawrence J Appel, Seth S Martin
2016-05-01
Abstract:Methods| A Johns Hopkins University School of Medicine institutional review board approved this study. In August and September 2015, we enrolled patients and staff who were at least 18 years old from 5 ambulatory Johns Hopkins sites (1 clinic each in general internal medicine, nephrology, and the Pro-Health clinical research unit, and 2 in cardiology clinics). Clinicians referred patients with or without hypertension or a baseline hypertensive reading for enrollment. After prescreening 105 individuals, written informed consent was obtained from 101 participants. Participants were given $5 gift cards for their time. Per prespecified dropping rules, data from 16 individuals were discarded because of unavailable cuff sizes (n= 1), standard device errors (n= 2), and excessive variation in sequential standard BP measurements, greater than 12 mm Hg for systolic BP or greater than 8 mm Hg for diastolic BP (n= 13). 3 For …