A device for reducing the atmosphere-induced interferences for analysis using inductively coupled plasma-mass spectrometry
Xin Jiang,Yongsheng Liu,Wengui Liu,Jie Lin,Zhenyi Liu,Lifei Chen,Xi Zhu,Wen Zhang,Zhaochu Hu
DOI: https://doi.org/10.1039/d4ja00185k
2024-08-18
Journal of Analytical Atomic Spectrometry
Abstract:The atmosphere-induced interference reduction device (AIRD) shows promise in mitigating interferences and enhancing elemental sensitivity in inductively coupled plasma-mass spectrometry (ICP-MS) analyses. In this study, the effectiveness of AIRD in reducing interferences caused by atmospheric gases such as C, N, O, ArN and ArO was evaluated. Argon (Ar) and helium (He) were chosen as shielding gases due to their inert properties. Krypton (Kr) served as the internal standard for monitoring instrumental conditions and correcting sensitivity drift. Gas dynamics simulations optimized the shielding gas flow rate, revealing a threshold of 10 L min-1 for optimal interference reduction. Results indicated the superiority of He in reducing interferences, with a reduction of 32% for 12C+, 51% for 15N+, 56% for 16O+, 64% for 17O+, 54% for 18O+, 51% for 54ArN+, and 42% for 56ArO+ compared to Ar. Moreover, AIRD maintained low oxide yields even after shielding gas cessation, with oxide yield maintained at approximately 0.03% for 24 hours. Experimental investigations further revealed that AIRD enhanced elemental sensitivity, particularly with He as the shielding gas, with a ~25% enhancement observed in the signal intensity of Kr.
spectroscopy,chemistry, analytical