Estimating the location of the hook-wire used in breast-conserving surgery using a magnetometer

Oiendrila Bhowmik Debnath,Akihiro Kuwahata,Yuki Sunaga,Shinichi Chikaki,Miki Kaneko,Moriaki Kusakabe,Masaki Sekino
DOI: https://doi.org/10.1063/9.0000799
IF: 1.697
2024-01-01
AIP Advances
Abstract:In the surgical treatment of nonpalpable breast lesions, such as in early-stage cancer, a hook-wire is inserted into the lesion as a marker to enable surgeons to excise the tissue, along with the hook-wire, with a good margin. However, a benchmark technique for intraoperatively determining whether the excised tissue has an appropriate margin around the lesion has not yet been established. In this study, a method for locating a ferromagnetic stainless steel hook-wire inside the excised tissue using a magnetometer is proposed. The magnetometer is placed around a phantom along with the hook-wire at varied locations to map the magnetic field distribution. The three-dimensional coordinates of hook-wire are obtained by executing an optimization algorithm. The experimental results indicate that the location of the hook-wire is successfully obtained. Based on the information regarding the margin around the hook-wire, the surgeon can immediately evaluate the risk of whether some cancer cells still remain in the body.
materials science, multidisciplinary,physics, applied,nanoscience & nanotechnology
What problem does this paper attempt to address?