Peripheral blood to next-generation sequencing ready DNA library: a novel engineering design for automation

Dulguunnaran Naranbat,Lothar À Brassard,Nabil Lawandy,Anubhav Tripathi
DOI: https://doi.org/10.1186/s12864-024-10892-0
IF: 4.547
2024-10-22
BMC Genomics
Abstract:Whole genome sequencing (WGS) has become a gold standard for diagnosing genomic variation. Peripheral blood is a common sample source for the extraction of nucleic acids for Next-Generation Sequencing (NGS) applications. Here, we present an integrated and fully automated device design that uses new concepts of fluid mechanics, heat-mass transfer, and thermodynamics of enzymatic reactions to extract nucleic acids from the blood and perform DNA library preparation from a pre-filled plate. We demonstrate that the presented device effectively extracts dsDNA with an average of 25.03 µg/mL and 25.91 µg/mL yield from citrate-stabilized human peripheral blood stored in Fresh (4 °C) and Frozen (-20 °C) conditions, respectively. Furthermore, our method automatically extracts nucleic acids and creates a high-quality sequence-ready DNA library from blood stabilized with citrate and EDTA for 8 samples simultaneously in a single run with a total operation time of ~ 7 h. Our results show the required coverage and depth of the genome, highlighting an essential application of this device in processing blood samples for genome sequencing.
What problem does this paper attempt to address?