VALIDATION OF QUANTITATIVE ELISAs FOR MEASURING ANTI-YERSINIA PESTIS F1 AND V ANTIBODY CONCENTRATIONS IN NONHUMAN PRIMATE SERA

C. Lindsey,B. Powell,C. Bolt,J. Brown,J. Adamovicz
DOI: https://doi.org/10.1080/15321819.2011.603774
2012-01-01
Abstract:This study systematically validated two quantitative enzyme-linked immunosorbent assays (ELISAs) for determining Yersinia pestis anti-F1 or anti-V IgG concentration in cynomolgus macaque sera. The results demonstrated that these ELISAs are reliable, reproducible, and suitable for their intended use to measure both anti-F1 and anti-V IgG in monkey sera following vaccination with a heterologous recombinant fusion F1-V protein (rF1-V). Statistical analysis demonstrated assay precision, accuracy, specificity, linearity/dilutional linearity, and robustness for both assays. The quantitative ranges of standard curves were defined as 40–700 ng/mLfor both anti-F1 and anti-V IgG. Either serological assay could be used to determine potency of F1/V antigen-based vaccines in surrogate clinical studies or to define correlates of protective immunity against plague under the Food and Drug Administration's (FDA) two-animal rule.
What problem does this paper attempt to address?