Evaluation of the EUROIMMUN automated chemiluminescence immunoassays for measurement of four core biomarkers for Alzheimer's disease in cerebrospinal fluid

Katharina Römpler,Philipp Arendt,Britta Brix,Viola Borchardt-Lohölter,Anette Schulz,Mandy Busse,Stefan Busse
DOI: https://doi.org/10.1016/j.plabm.2024.e00425
2024-09-05
Abstract:Introduction: Robust immunoassays for quantification of Alzheimer's disease (AD)-specific biomarkers are required for routine diagnostics. We report analytical performance characteristics of four new chemiluminescence immunoassays (ChLIA, EUROIMMUN) running on closed, fully automated random-access instruments for quantification of Aβ1-40, Aβ1-42, tTau, and pTau(181) in human cerebrospinal fluid (CSF). Methods: ChLIAs were validated according to the guidelines of the Clinical and Laboratory Standards Institute (CLSI). Optimal cut-offs for biomarkers and biomarker ratios were determined using samples from 219 AD patients and 220 patients with AD-related symptoms. For performance comparison, biomarker concentrations were measured in 110 diagnostic leftover samples using the ChLIAs and established Lumipulse G assays (Fujirebio). Results: All ChLIAs met CLSI criteria. Overall agreement between assays was 89.0%-97.3 % with highly correlating results (Pearson's correlation coefficients: 0.82-0.99). Passing-Bablok regression analysis revealed systematic differences. Discussion: EUROIMMUN ChLIAs showed good analytical performances and represent new valuable tools for diagnostics of AD.
What problem does this paper attempt to address?