CSF Biomarkers in Longitudinal Alzheimer Disease Cohorts: Pre-Analytic Challenges

Erin M Jonaitis,Beckie Jeffers,Monica VandenLangenberg,Yue Ma,Carol Van Hulle,Rebecca Langhough,Lianlian Du,Nathaniel A Chin,Robert J Przybelski,Kirk J Hogan,Bradley T Christian,Tobey J Betthauser,Ozioma C Okonkwo,Barbara B Bendlin,Sanjay Asthana,Cynthia M Carlsson,Sterling C Johnson
DOI: https://doi.org/10.1093/clinchem/hvad221
IF: 12.114
2024-03-01
Clinical Chemistry
Abstract:Abstract Background The sensitivity of amyloid to pre-analytic factors complicates cerebrospinal fluid (CSF) diagnostics for Alzheimer disease. We report reliability and validity evidence for automated immunoassays from frozen and fresh CSF samples in an ongoing, single-site research program. Methods CSF samples were obtained from 2 Wisconsin cohorts (1256 measurements; 727 participants). Levels of amyloid beta 1–42 (Aβ42), phosphorylated tau 181 (pTau181), and total tau (tTau) were obtained using an Elecsys cobas e 601 platform. Repeatability and fixed effects of storage tube type, extraction method, and freezing were assessed via mixed models. Concordance with amyloid positron emission tomography (PET) was investigated with 238 participants having a temporally proximal PET scan. Results Repeatability was high with intraclass correlation (ICC) ≥0.9, but tube type strongly affected measurements. Discriminative accuracy for PET amyloid positivity was strong across tube types (area under the curve [AUC]: Aβ42, 0.87; pTau181Aβ42 , 0.96), although optimal thresholds differed. Conclusions Under real-world conditions, the Elecsys platform had high repeatability. However, strong effects of pre-analytic factors suggest caution in drawing longitudinal inferences.
medical laboratory technology
What problem does this paper attempt to address?