Modified Beer-Lambert algorithm to measure pulsatile blood flow, critical closing pressure, and intracranial hypertension

Wesley B Baker,Rodrigo M Forti,Pascal Heye,Kristina Heye,Jennifer M Lynch,Arjun G Yodh,Daniel J Licht,Brian R White,Misun Hwang,Tiffany S Ko,Todd J Kilbaugh
DOI: https://doi.org/10.1364/BOE.529150
2024-08-27
Abstract:We introduce a frequency-domain modified Beer-Lambert algorithm for diffuse correlation spectroscopy to non-invasively measure flow pulsatility and thus critical closing pressure (CrCP). Using the same optical measurements, CrCP was obtained with the new algorithm and with traditional nonlinear diffusion fitting. Results were compared to invasive determination of intracranial pressure (ICP) in piglets (n = 18). The new algorithm better predicted ICP elevations; the area under curve (AUC) from logistic regression analysis was 0.85 for ICP ≥ 20 mmHg. The corresponding AUC for traditional analysis was 0.60. Improved diagnostic performance likely results from better filtering of extra-cerebral tissue contamination and measurement noise.
What problem does this paper attempt to address?