All-optics technique for monitoring absolute cerebral blood flow: validation against magnetic resonance imaging perfusion

Leena N Shoemaker,Saeed Samaei,Graham Deller,Danny J J Wang,Daniel Milej,Keith St Lawrence
DOI: https://doi.org/10.1117/1.NPh.11.4.045002
Neurophotonics
Abstract:Significance: The ability to monitor cerebral blood flow (CBF) at the bedside is essential to managing critical-care patients with neurological emergencies. Diffuse correlation spectroscopy (DCS) is ideal because it is non-invasive, portable, and inexpensive. We investigated a near-infrared spectroscopy (NIRS) approach for converting DCS measurements into physiological units of blood flow. Aim: Using magnetic resonance imaging perfusion as a reference, we investigated the accuracy of absolute CBF measurements from a bolus-tracking NIRS method that used transient hypoxia as a flow tracer and hypercapnia-induced increases in CBF measured by DCS. Approach: Twelve participants (7 female, 28 ± 6 years) completed a hypercapnia protocol with simultaneous CBF recordings from DCS and arterial spin labeling (ASL). Nine participants completed the transient hypoxia protocol while instrumented with time-resolved NIRS. The estimate of baseline CBF was subsequently used to calibrate hypercapnic DCS data. Results: Moderately strong correlations at baseline ( slope = 0.79 and R 2 = 0.59 ) and during hypercapnia ( slope = 0.90 and R 2 = 0.58 ) were found between CBF values from calibrated DCS and ASL (range 34 to 85 mL / 100 g / min ). Conclusions: Results demonstrated the feasibility of an all-optics approach that can both quantify CBF and perform continuous perfusion monitoring.
What problem does this paper attempt to address?