The first clinical implementation of real-time 6 degree-of-freedom image-guided radiotherapy for liver SABR patients
Chandrima Sengupta,Doan Trang Nguyen,Trevor Moodie,Daniel Mason,Jianjie Luo,Trent Causer,Sau Fan Liu,Elizabeth Brown,Lauren Inskip,Maryam Hazem,Menglei Chao,Tim Wang,Yoo Y Lee,Kirsten van Gysen,Emma Sullivan,Eireann Cosgriff,Prabhakar Ramachandran,Per Poulsen,Jeremy Booth,Ricky O'Brien,Peter Greer,Paul Keall,Yoo Y. Lee
DOI: https://doi.org/10.1016/j.radonc.2023.110031
IF: 6.901
2024-01-01
Radiotherapy and Oncology
Abstract:PURPOSE: Multiple survey results have identified a demand for improved motion management for liver cancer IGRT. Until now, real-time IGRT for liver has been the domain of dedicated and expensive cancer radiotherapy systems. The purpose of this study was to clinically implement and characterise the performance of a novel real-time 6 degree-of-freedom (DoF) IGRT system, Kilovoltage Intrafraction Monitoring (KIM) for liver SABR patients.METHODS/MATERIALS: The KIM technology segmented gold fiducial markers in intra-fraction x-ray images as a surrogate for the liver tumour and converted the 2D segmented marker positions into a real-time 6DoF tumour position. Fifteen liver SABR patients were recruited and treated with KIM combined with external surrogate guidance at three radiotherapy centres in the TROG 17.03 LARK multi-institutional prospective clinical trial. Patients were either treated in breath-hold or in free breathing using the gating method. The KIM localisation accuracy and dosimetric accuracy achieved with KIM + external surrogate were measured and the results were compared to those with the estimated external surrogate alone.RESULTS: The KIM localisation accuracy was 0.2±0.9 mm (left-right), 0.3±0.6 mm (superior-inferior) and 1.2±0.8 mm (anterior-posterior) for translations and -0.1<sup>◦</sup>±0.8<sup>◦</sup> (left-right), 0.6<sup>◦</sup>±1.2<sup>◦</sup> (superior-inferior) and 0.1<sup>◦</sup>±0.9<sup>◦</sup> (anterior-posterior) for rotations. The cumulative dose to the GTV with KIM + external surrogate was always within 5% of the plan. In 2 out of 15 patients, >5% dose error would have occurred to the GTV and an organ-at-risk with external surrogate alone.CONCLUSIONS: This work demonstrates that real-time 6DoF IGRT for liver can be implemented on standard radiotherapy systems to improve treatment accuracy and safety. The observations made during the treatments highlight the potential false assurance of using traditional external surrogates to assess tumour motion in patients and the need for ongoing improvement of IGRT technologies.
oncology,radiology, nuclear medicine & medical imaging