A 3-Dimensional finite element analysis of patient-specific implant (with strut abutments) interface on stress concentration on the implant and surrounding bone in bilateral maxillary deformities

Abhay N Datarkar,Arwa F Pardiwala,Priyanka Relan,Surendra Daware,Vandana Gadve,Archana Deshpande,Ashlesha Ghormade
DOI: https://doi.org/10.4103/njms.njms_155_23
Abstract:Purpose: To determine the efficacy and longevity of patient-specific implants (PSIs) with strut abutment design to rehabilitate bilateral maxillectomy defect. Materials and methods: Finite Element Analysis was performed on a PSI with strut abutments to repair a patient's bilateral maxillectomy defect due to COVID associated mucormycosis. Results: The von Mises stress recorded was maximum in the zygomaticomaxillary buttress region, and displacement values were noted to be highest in the posterior-most strut, although both parameters were within acceptable limits, which is favorable. Conclusion: The authors draw the conclusion that a PSI with strut abutments is a workable therapeutic modality for patients with these kinds of abnormalities based on this information.