Bailout Procedure Utilizing Balloon Dilatation for a Percutaneous Micro-axial Flow Pump Entrapped Within a Significantly Calcified Subclavian Artery

Kazuyoshi Takagi,Hiroyuki Otsuka,Kosuke Saku,Eiki Tayama
DOI: https://doi.org/10.7759/cureus.65804
2024-07-30
Cureus
Abstract:The IMPELLA 5.5 (Abiomed Inc., Danvers, Massachusetts, United States) is a catheter-based, micro-axial blood pump designed to enhance organ perfusion in patients with cardiogenic shock. Despite its superior hemodynamic support, vascular complications are a significant concern, with many patients needing to discontinue IMPELLA therapy due to these issues. Patients may even require surgical intervention to address device-related vascular injuries. The IMPELLA 5.5 implantation in vessels with severe calcification is particularly associated with complications such as vascular calcification, stenosis, vascular tortuosity, and the use of larger sheaths are risk factors following endovascular therapy and IMPELLA implantation. In this report, we present a case of severe calcification in the right subclavian artery, in which the IMPELLA 5.5 was lodged. The calcifications protruded into the vascular lumen, becoming lodged between the IMPELLA motor and the cannula, complicating extraction despite the vessel having sufficient diameter. We successfully removed the device using a balloon dilation technique, ensuring safe extraction. No vascular complications such as pseudoaneurysm or dissection were observed in the right subclavian artery one month after extraction. This case highlights a potential approach for managing similar complications and vascular access for IMPELLA insertion.