Assessment of accuracy and repeatability of quantitative parameter mapping in MRI

Yuya Hirano,Kinya Ishizaka,Hiroyuki Sugimori,Yo Taniguchi,Tomoki Amemiya,Yoshitaka Bito,Kohsuke Kudo
DOI: https://doi.org/10.1007/s12194-024-00836-4
2024-08-28
Abstract:We aimed to evaluate the accuracy and repeatability of the T1, T2*, and proton density (PD) values obtained by quantitative parameter mapping (QPM) using the ISMRM/NIST MRI system phantom and compared them with computer simulations. We compared the relaxation times and PD obtained through QPM with the reference values of the ISMRM/NIST MRI system phantom and conventional methods. Furthermore, we evaluated the presence or absence of influences other than noise in T1 and T2* values obtained by QPM by comparing the obtained coefficient of variation (CV) with simulation results. The T1, T2*, and PD values by QPM showed a strong correlation with the measured values and the referenced values. The simulated CVs of QPM calculated for each sphere showed similar trends to those of the actual scans.