In vitro heat insulation efficacy of 5% dextrose versus 0.9% saline during radiofrequency ablation

Yanping Ma,Jinfen Wang,Tao Wu,Bowen Zheng,Tinghui Yin,Yufan Lian,Jie Ren
DOI: https://doi.org/10.14366/usg.24073
2024-07-15
ULTRASONOGRAPHY
Abstract:Purpose: This study compared the efficacy of heat insulation between 5% dextrose and 0.9% saline in radiofrequency ablation (RFA). Accordingly, temperature variations and maximum temperatures were assessed at identical distances and heat field distributions. Methods: Cubes of porcine liver tissue, measuring 10 mm across, were selected to precisely align the ablation boundary with the tissue boundary. An 18-gauge electrode with a 7-mm tip was inserted into each cube (10 per group) in a stainless-steel cup containing 40 mL of 5% dextrose or 0.9% saline. Fixed ablation was performed for 3 minutes using continuous mode at 30 W, simulating the typical thermal environment during thyroid RFA. Real-time temperature measurements were recorded by sensors positioned 0, 1, 3, and 5 mm from the cube's edge. A comparative analysis was conducted to assess the maximum temperature, temperature variation, and duration of temperatures exceeding 42℃. Results: In both groups, the temperature curve declined with increasing distance from the edge of the ablated tissue. However, 0.9% saline exhibited higher maximum temperatures at 1, 3, and 5 mm compared to 5% dextrose (1 mm: 44.55°C±5.25°C vs. 34.68°C±3.07°C; 3 mm: 39.64°C±2.53°C vs. 29.22°C±2.21°C; 5 mm: 38.86°C±2.14°C vs. 28.74°C±2.51°C; all P<0.001). Considering a nerve injury threshold of 42°C, the 0.9% saline also displayed a greater proportion of samples reaching this temperature and a longer duration of temperatures exceeding it (P<0.05). Conclusion: The heat insulation efficacy of 5% dextrose at 1-5 mm exceeds that of 0.9% saline at identical distances and in a common thermal environment during thyroid RFA.
What problem does this paper attempt to address?