MPI System with Bore Sizes of 75 mm and 100 mm Using Permanent Magnets and FMMD Technique

Jae Chan Jeong,Tae Yi Kim,Hyeon Sung Cho,Beom Su Seo,Hans Joachim Krause,Hyo Bong Hong
DOI: https://doi.org/10.3390/s24123776
IF: 3.9
2024-06-11
Sensors
Abstract:We present two magnetic particle imaging (MPI) systems with bore sizes of 75 mm and 100 mm, respectively, using three-dimensionally arranged permanent magnets for excitation and frequency mixing magnetic detection (FMMD) coils for detection. A rotational and a translational stage were combined to move the field free line (FFL) and acquire the MPI signal, thereby enabling simultaneous overall translation and rotational movement. With this concept, the complex coil system used in many MPI systems, with its high energy consumption to generate the drive field, can be replaced. The characteristic signal of superparamagnetic iron oxide (SPIO) nanoparticles was generated via movement of the FFL and acquired using the FMMD coil. The positions of the stages and the occurrence of the f1 + 2f2 harmonics were mapped to reconstruct the spatial location of the SPIO. Image reconstruction was performed using Radon and inverse Radon transformations. As a result, the presented method based on mechanical movement of permanent magnets can be used to measure the MPI, even for samples as large as 100 mm. Our research could pave the way for further technological developments to make the equipment human size, which is one of the ultimate goals of MPI.
engineering, electrical & electronic,chemistry, analytical,instruments & instrumentation
What problem does this paper attempt to address?
The main problem that this paper attempts to solve is to develop a magnetic particle imaging (MPI) system using permanent magnets and frequency - mixed magnetic detection (FMMD) technology in order to measure larger samples (such as samples with a diameter of up to 100 mm). Specifically, the research aims to generate a field - free line (FFL) by mechanically moving the permanent magnets, thereby replacing the traditional driving - field generation method, reducing the energy consumption of the device, and simplifying the system structure. In addition, this study also explores how to use this system for high - resolution two - dimensional and three - dimensional imaging, paving the way for the future application of MPI technology to human - scale devices. ### Research Background - **MPI Technology**: Magnetic particle imaging (MPI) is a medical imaging technology that uses the nonlinear magnetic response characteristics of superparamagnetic iron oxide (SPIO) nanoparticles for imaging. Unlike MRI, MPI does not provide the same anatomical information, but has the advantages of high resolution and fast acquisition speed. - **Existing Challenges**: Existing MPI systems usually require complex coil systems to generate driving fields, which leads to high energy consumption and device complexity. In addition, the maximum aperture of current MPI systems is only 20 cm, which limits its application to large animals or humans. ### Research Objectives - **Develop a New MPI System**: Use permanent magnets and FMMD technology to design and implement MPI systems with two different apertures (75 mm and 100 mm). - **Mechanically Move FFL**: Generate FFL by rotating and translating the permanent magnets through mechanical means, thereby replacing the traditional driving - field generation method. - **High - Resolution Imaging**: Verify the sensitivity and spatial resolution of the system and conduct two - dimensional and three - dimensional imaging experiments. ### Technical Details - **Permanent Magnet Arrangement**: Use 24 NdFeB (N35 grade) permanent magnets, arranged in two overlapping aluminum housings to generate a selection field. - **FMMD Coil**: Used to detect the signals of SPIO nanoparticles. The coil system consists of two excitation coils and one detection coil. - **Signal Processing**: Use a DAQ system to process the signals measured by the detection coil and perform image reconstruction through Radon and inverse Radon transforms. ### Experimental Results - **Sensitivity**: The system can detect SPIO concentrations as low as 0.125 mg/mL, and the detection results are consistent at different positions (center and edge). - **Spatial Resolution**: The spatial resolution was tested by placing three PCR tubes, and the results showed that the system can clearly distinguish the SPIO distribution at different positions. - **Two - Dimensional and Three - Dimensional Imaging**: Successfully obtained two - dimensional and three - dimensional MPI images, verifying the imaging ability of the system. ### Significance - **Reduce Energy Consumption**: Significantly reduce the energy consumption of the system by mechanically moving the permanent magnets to generate FFL. - **Simplify the Structure**: Reduce the complex coil system in the traditional MPI system, making the device more compact and easier to operate. - **Expand the Scope of Application**: Provide the possibility for the future application of MPI technology to human - scale devices, and are expected to play an important role in the field of medical imaging. In conclusion, this study has developed an efficient, low - energy - consumption MPI system suitable for larger samples through innovative permanent magnet and FMMD technologies, laying the foundation for the future development of medical imaging technology.