4D printing of overall radiopaque customized bionic occlusion devices
Cheng Lin,Zhipeng Huang,Qinglong Wang,Zhichen Zou,Wenbo Wang,Liwu Liu,Yanju Liu,Jinsong Leng
DOI: https://doi.org/10.1002/adhm.202201999
IF: 10
2022-11-08
Advanced Healthcare Materials
Abstract:Percutaneous closure of ventricular septal defect (VSD) can effectively occlude abnormal blood flow between ventricles. However, commonly used Nitinol occlusion devices have non‐negligible limitations, such as non‐degradability leading to life‐threatening embolization; limited device size predisposing to displacement and wear; only a few radiopaque markers resulting in inaccurate positioning. Nevertheless, the exploration of customized, biodegradable, and overall radiopaque occluders is still vacant. Here, overall radiopaque, biodegradable, and dynamic reconfigurable 4D printed VSD occluders were developed. Based on wavy bionic structures, various VSD occluders were designed and manufactured to adapt to the position diversity of VSD. The customized configuration, biocompatibility, and biodegradability of the developed 4D printed bionic occluders can eliminate the series of complications caused by traditional occluders. The overall radiopacity of 4D printed VSD occluders was validated ex vivo and in vivo, whereby accurate positioning can be assured. Notably, the preparation strategies for 4D printed occluders are scalable, eliminating the barriers to mass production, and marking a meaningful step in bridging the gap between modeling and clinical application of 4D printed occlusion devices. This work opens attractive perspectives for the rapid manufacturing of customized intelligent medical devices for which overall radiopacity, dynamic reconfigurability, biocompatibility, and biodegradability are sought. This article is protected by copyright. All rights reserved
nanoscience & nanotechnology,engineering, biomedical,materials science, biomaterials