Operating Room Air May Harbor Pathogens: The Role of an Ultraviolet Air Filtration Unit

Diana Fernández-Rodríguez,Saad Tarabichi,Krystal Golankiewicz,Nicolina Zappley,Javad Parvizi
DOI: https://doi.org/10.1016/j.arth.2024.05.072
Abstract:Background: One important factor for the prevention of surgical site infections is ultraclean air in the operating room (OR). Still, the direct sterilization potential of most technologies, especially in a dynamic clinical setting, is not well understood. We aimed to determine and compare the microbial presence from the inlet and outlet flow of a filtration unit with crystalline ultraviolet-C (C-UVC) light. Methods: A prospective study was conducted at a single institution, where primary total joint arthroplasty and spine surgeries were performed. The OR was fitted with a positive ventilation system. In addition, a filtration unit with a C-UVC sterilizing light was placed in the OR. The inlet and outlet air flows were swabbed simultaneously and compared. Swabs were processed for culture and next-generation sequencing. Results: The mean length of the surgical procedures sampled was 68 ± 13 minutes. Overall, 19 out of 200 (9.5%) swabs isolated microorganisms. Inlet air swabs were positive at a higher rate (16 versus 3%; P < .01) compared to the outlet air swabs. A wide variety of Gram-positive, Gram-negative, and anaerobic bacteria were isolated, but fungi were only recovered from inlet air swabs. The detection of microorganisms was also higher when more door openings were performed (32.5 ± 7.1 versus 27.9 ± 5.6; P < .01). Conclusions: Air swabs mainly isolated microorganisms from the inlet flow to the filtration unit with a C-UVC light. The sterilizing unit counteracted factors affecting the air quality in the OR, namely door openings, surgical personnel, and tissue combustion.
What problem does this paper attempt to address?