Pitfalls and tips for lumenless lead extraction inserted deep within the ventricular septum

Nobunari Tomura,Hirokazu Shiraishi,Masahiro Makino,Jun Munakata,Satoshi Shimoo,Hibiki Iwakoshi,Tetsuro Nishimura,Takashi Ohkura,Keitaro Senoo,Satoaki Matoba
DOI: https://doi.org/10.1002/ccr3.8718
2024-04-24
Abstract:Key clinical message: This case highlights the pitfalls and provides tips for the extraction of deeply implanted lumenless leads, and encourages careful lead selection in the current era of widespread left bundle branch area pacing. Abstract: The extraction of cardiovascular implantable electronic device leads is sometimes complicated. We describe a case with difficult but successful extraction of SelectSecure, a lumenless permanent pacemaker lead, implanted deep in the ventricular septum, highlighting its pitfalls and tips in the current era of left bundle branch area pacing.
What problem does this paper attempt to address?