Framewise multi-echo distortion correction for superior functional MRI

Andrew N Van,David F Montez,Timothy O Laumann,Vahdeta Suljic,Thomas Madison,Noah J Baden,Nadeshka Ramirez-Perez,Kristen M Scheidter,Julia S Monk,Forrest I Whiting,Babatunde Adeyemo,Roselyne J Chauvin,Samuel R Krimmel,Athanasia Metoki,Aishwarya Rajesh,Jarod L Roland,Taylor Salo,Anxu Wang,Kimberly B Weldon,Aristeidis Sotiras,Joshua S Shimony,Benjamin P Kay,Steven M Nelson,Brenden Tervo-Clemmens,Scott A Marek,Luca Vizioli,Essa Yacoub,Theodore D Satterthwaite,Evan M Gordon,Damien A Fair,M Dylan Tisdall,Nico U F Dosenbach
DOI: https://doi.org/10.1101/2023.11.28.568744
2023-11-29
bioRxiv
Abstract:Functional MRI (fMRI) data are severely distorted by magnetic field (B0) inhomogeneities which currently must be corrected using separately acquired field map data. However, changes in the head position of a scanning participant across fMRI frames can cause changes in the B0 field, preventing accurate correction of geometric distortions. Additionally, field maps can be corrupted by movement during their acquisition, preventing distortion correction altogether. In this study, we use phase information from multi-echo (ME) fMRI data to dynamically sample distortion due to fluctuating B0 field inhomogeneity across frames by acquiring multiple echoes during a single EPI readout. Our distortion correction approach, MEDIC (Multi-Echo DIstortion Correction), accurately estimates B0 related distortions for each frame of multi-echo fMRI data. Here, we demonstrate that MEDIC's framewise distortion correction produces improved alignment to anatomy and decreases the impact of head motion on resting-state functional connectivity (RSFC) maps, in higher motion data, when compared to the prior gold standard approach (i.e., TOPUP). Enhanced framewise distortion correction with MEDIC, without the requirement for field map collection, furthers the advantage of multi-echo over single-echo fMRI.
What problem does this paper attempt to address?