Dermal tissue penetration of in-plane silicon microneedles evaluated in skin-simulating hydrogel, rat skin and porcine skin

Stephanie Ingemann Bisgaard,Long Quang Nguyen,Katrine Lindholm Bøgh,Stephan Sylvest Keller
DOI: https://doi.org/10.1016/j.bioadv.2023.213659
Abstract:Recently, microneedle-based sensors have been introduced as novel strategy for in situ monitoring of biomarkers in the skin. Here, in-plane silicon microneedles with different dimensions and shapes are fabricated and their ability to penetrate skin is evaluated. Arrays with flat, triangular, hypodermic, lancet and pencil-shaped microneedles, with lengths of 500-1000 μm, widths of 200-400 μm and thickness of 180-500 μm are considered. Fracture force is higher than 20 N for all microneedle arrays (MNA) confirming a high mechanical stability of the microneedles. Penetration force in skin-simulating hydrogels, excised rat abdominal skin and porcine ear skin is at least five times lower than the fracture force for all MNA designs. The lowest force for skin penetration is required for triangular microneedles with a low width and thickness. Skin tissue staining and histological analysis of rat abdominal skin and porcine ear skin confirm successful penetration of the epidermis for all MNA designs. However, the penetration depth is between 100 and 300 μm, which is considerably lower than the microneedle length. Tissue damage estimated by visual analysis of the penetration hole is smallest for triangular microneedles. Penetration ability and tissue damage are compared to the skin prick test (SPT) needle applied in allergy testing.