Characterisation of a Sub-20 Ps Temporal Resolution Pulse Dilation Photomultiplier Tube.
S. G. Gales,C. J. Horsfield,A. L. Meadowcroft,A. E. Leatherland,H. W. Herrmann,J. D. Hares,A. K. L. Dymoke-Bradshaw,J. S. Milnes,Y. H. Kim,H. G. Kleinrath,K. Meaney,A. B. Zylstra,S. Parker,D. Hussey,L. Wilson,S. F. James,J. D. Kilkenny,T. J. Hilsabeck
DOI: https://doi.org/10.1063/1.5031110
IF: 1.6
2018-01-01
Review of Scientific Instruments
Abstract:A pulse-dilation photomultiplier tube (PD-PMT) with sub-20 ps temporal resolution has been developed for use with γ-ray-sensitive gas Cherenkov detectors at the National Ignition Facility to improve the diagnosis of nuclear fusion burn history and the areal density of the remaining capsule ablator. The pulse-dilation mechanism entails the application of a time-dependent, ramp waveform to a photocathode-mesh structure, introducing a time-dependent photoelectron accelerating potential. The electric field imparts axial velocity dispersion to outgoing photoelectrons. The photoelectron pulse is dilated as it transits a drift region prior to amplification in a microchannel plate and read out with a digital oscilloscope. We report the first measurements with the prototype PD-PMT demonstrating nominal <20 ps FWHM across a 400 ps measurement window and <30 ps FWHM for an extracted charge up to 300 pC. The output peak areas are linear to within 20% over 3 orders of magnitude of input intensity. 3D particle in cell simulations, which included space charge effects, have been carried out to investigate the device temporal magnification, resolution, and linearity.