Mechanical characterisation of hydrophobic and hydrophilic acrylates used in intraocular lenses through depth sensing indentation

I Cabeza-Gil,B Calvo,A Rico,C Reinhards-Hervás,J Rodríguez
DOI: https://doi.org/10.1016/j.jmbbm.2021.104997
Abstract:In this work, the mechanical behaviour of hydrophilic and hydrophobic acrylates has been characterised by depth sensing indentation. Time-dependent behaviour has been studied using load-relaxation tests. Experiments have been simulated with a finite element software using a visco-hyperelastic material model. The parameters of this model have been determined using deep learning techniques. The developed material models have been used to mechanically simulate a standard compression test of a prototype intraocular lens.
What problem does this paper attempt to address?