Orthogonal Optimization Design of Structural Parameters for Bioaerosol Sampler Using Computational Fluid Dynamics Simulation and Field Experiments
Xuezheng Ma,Zhijian Liu,Fanshaung Li,Yu Niu,Beibei Wang,Huipeng Li,Haiyang Liu,Rui Rong,Guoyan Wang,Liping Zhang,Jinsong Li,Kongxin Hu,Tao Zheng
DOI: https://doi.org/10.1080/02786826.2020.1813873
2020-01-01
Aerosol Science and Technology
Abstract:Aerosol-to-hydrosol biological samplers have a low particle collection rate and aerosol-to-liquid efficiency owing to their airflow rate and geometry. The study develops a method to optimize the design of an aerosol-to-hydrosol bioaerosol sampler for better particle-collection efficiency. Four important performance parameters-particle size, nozzle angle, liquid container shape, and distance between nozzle and liquid surface-were optimized using the orthogonal design method. These parameters, along with the particle-capturing characteristics, were predicted using an approach based on computational fluid dynamics. Subsequently, this approach was used to test the bioaerosol's capture efficiency. A prototype based on the optimized geometrical design was 3D-printed and tested in a field experiment. The most optimal parameters were particle size = 10 mu m; inverted cone shape; liquid-level-nozzle distance = 10 mm; and angle = 45 degrees. Copyright (c) 2020 American Association for Aerosol Research