Monopolar Radiofrequency Energy Delivered by a Conductive Endovascular Basket or Guidewire Leads to Thermal Occlusion in a Swine Model

John W Karanian,Venkatesh P Krishnasamy,Russ M Guidry,Rachel Hecht,David L Woods,Juan A Esparza-Trujillo,Ivane Bakhutshvili,William F Pritchard,Bradford J Wood
DOI: https://doi.org/10.1016/j.jvir.2020.04.035
Abstract:Purpose: To assess the feasibility of inducing vascular occlusion by application of radiofrequency (RF) energy via conductive endovascular wires or baskets. Materials and methods: A retrievable nitinol basket and stainless steel guidewire with a platinum tip were evaluated as conductors for endovascular application of RF energy. Tissue-mimicking thermochromic gel phantoms that change color with heating were cast with 2-, 5-, and 7-mm-diameter lumens and filled with 37 oC saline. After ablation, the phantoms were sectioned, and the thermal footprints were evaluated. Six castrated male domestic swine underwent endovascular ablation using the basket in iliac arteries and guidewires in renal arteries. Post-procedural angiography was performed, and postmortem arterial segments were resected for histopathologic analysis. Results: In the phantom, the depth of thermal change in the 5- and 7-mm lumens averaged 6.3 and 6.0 mm along the basket, respectively, and in the 2- and 5-mm lumens, the depth of thermal change averaged 1.9 and 0.5 mm along the wire, respectively. In the swine, RF energy delivery led to angiographic occlusion at 12 of 13 sites. Thermal injury and occlusion were similar at the proximal, middle, and distal basket treatment zone, whereas injury and occlusion decreased from the proximal to the distal end of the 5-cm wire treatment zone. Conclusions: Endovascular delivery of RF energy via a conductive basket in medium-sized arteries or a guidewire in small arteries led to acute angiographic and histologic occlusion. The potential to induce stasis might be useful in settings where rapid occlusion is desirable.
What problem does this paper attempt to address?