Rapid Deployment Valve Implantation using the EDWARDS Intuity Valve System: A Word of Caution regarding Sizing in Calcified Sinotubular Junctions

T. Wahlers,P. Rahmanian
DOI: https://doi.org/10.1055/s-0035-1546298
2015-03-10
The Thoracic and Cardiovascular Surgeon
Abstract:Abstract In aortic valve replacement, severe calcification of the sinotubular junction is a limitation in the application of the rapid deployment Edwards Intuity valve system. This is illustrated by a case presentation in which discrepancies of shape and diameter between sizer and valve mounted on the deployment system resulted in discarding the prosthesis and performing a biocomposite root replacement instead. Modification of the valve sizer will allow for safer sizing because the sizing process will mimic the implantation process more accurately.