A Signal-to-Cutoff Ratio in the Abbott Architect HIV Ag/Ab Combo Assay That Predicts Subsequent Confirmation of HIV-1 Infection in a Low-Prevalence Setting

T. Jensen,P. Robertson,R. Whybin,Ian Chambers,Monica Lahra,W. Rawlinson,J. Post
DOI: https://doi.org/10.1128/JCM.03583-14
2015-02-11
Journal of Clinical Microbiology
Abstract:ABSTRACT A rapid diagnosis is considered important in HIV care. In 138,911 testing episodes with the Abbott Architect HIV Ag/Ab Combo assay (3,705 reactive samples), a signal-to-cutoff ratio of >151.17 had a positive predictive value of 100% and a sensitivity of 67.4% for the detection of subsequently confirmed HIV infection. We suggest that results higher than this signal-to-cutoff ratio threshold may be reported to clinicians before the completion of confirmatory testing.