Non-matrix-matched calibration in bulk multi-element laser ablation – Inductively coupled plasma – Mass spectrometry analysis of diverse materials

Kristina Mervič,Vid S. Šelih,Martin Šala,Johannes T. van Elteren
DOI: https://doi.org/10.1016/j.talanta.2024.125712
IF: 6.1
2024-02-03
Talanta
Abstract:Laser ablation inductively coupled plasma - mass spectrometry (LA-ICP-MS) is a frequently used microanalytical technique in elemental analysis of solid samples. In most instances the use of matrix-matched calibration standards is necessary for the accurate determination of elemental concentrations. However, the main drawback of this approach is the limited availability of certified reference materials. Here, we present a novel conceptual framework in LA-ICP-MS quantification without the use of matrix-matched calibration standards but instead employment of an ablation volume-normalization method (via measurement of post-ablation line scan volumes by optical profilometry) in combination with a matrix-adapted fluence (slightly above the ablation threshold). This method was validated by cross-matrix quantification of reference materials typically investigated by LA-ICP-MS, including geological and biological materials. This allows for more accurate and precise multi-element quantification, and enables quantification of previously unquantifiable elements/materials.
chemistry, analytical
What problem does this paper attempt to address?